Sterilization gains (inline)
Procedure:
Valve (1) and valve (4) closes.
Drain valve (2) opens.
Valve (3) opens and maintain steam flow into the filter housing.
After reaching a temperature of 100 ° C, the steam begins to condense at the same time that there is only opened to the valve (2), the pressure being built up to the desired sterilization temperature.
After reaching the steam temperature starts the actual sterilization within the ages:
• Saturated steam 121 ° C - 30 minutes
• Saturated steam 131 ° C - 20 minutes
• Saturated steam 141 ° C - 10 minutes
When sterilization rounded cast of valve (2), after which valve (3) & (1) open slowly and valve (4) closes slowly - and then start the process over again.
Sterilization progress (item removed)
Be given to the sterilization of the filter element in a predetermined cycle, it may be every 14 days, 1 time per. month or every quarter
Ultra Filter recommend to be at least 1 time in quarter
It is often advantageous to have two elements, one is in the filter housing while the other is in the process of sterilization.
|